Juan J. AlióI; Miguel ClarosII; Eva M. RamírezIII; Carmen IglesiasIV; José Fernández-LlebrezV

Abstract

Objective: The purpose of this study was to research the in vitro behavior of JAL 90458 as a buffering agent which reduces friction between brackets and stainless steel arch wires of different cross sections and sizes.

Methods: Three types of stainless steel wires with different cross sections and three types of ligatures were used with and without JAL 90458 to measure the friction according to the time and distance traveled by the brackets (Sinergy®) in an inclined plane with two angulations. The Kruskal-Wallis one-way analysis of variance by ranks was applied to determine the degree of friction between the group using and the group not using the product
(P ≤ .05).

Results: Separate analysis of the arch wires, ligatures and angulation with and without compound revealed statistically significant differences between the groups, showing that friction was reduced significantly when the JAL 90458 compound was used. The 0.021x0.025" archwires and the arch wires attached using elastic ligatures produce the least resistance to sliding among all of those analyzed when the product was not used.

Conclusions: The results show that JAL 90458 reduces friction independently of the arch wire cross section, type of ligature and angulation of the measuring instrument.

Key words: Friction, JAL 90458, arch wires, ligatures, in vitro.

Introduction

Throughout orthodontic treatment, brackets move along the archwire or the archwire moves along tubes and brackets in the alignment and gap-closing stages using a sliding mechanism. During this movement, a negative physical phenomenon occurs which opposes free movement over surfaces and is known as friction or resistance to sliding. This is a deciding factor in treatment length.1-5 Friction is a clinical challenge, particularly in the sliding mechanism, and is the result of physical and biological parameters.6 Friction is also produced by using elastic pieces or metal ligatures that tie the arch to the bracket.7 It can also depend on other factors, such as surface texture (stick-slip phenomenon),8,9 the friction coefficient,10 the angulation formed between the axis of the wire and the bracket slot5,11,12,13 and the degree of plastic and elastic deformation of the wire (binding and notching)…etc.14,15

Friction increases the forces needed to move a tooth, causing periodontal damage, brackets becoming dislodged and loss of anchorage.5,19,22 When we move a tooth in the mesial or distal direction, unwanted rotations or angulations of the teeth are produced which can reduce the applied force by more than 50% due to friction.19 Studies show that friction increases as arch section and strength increase and it will be related to the type of ligation used.5,6,32 The arch wire-slot angulation forms a critical contact angle, bringing about a passive or active configuration according to the final degree of angulation produced. The greater the final angulation is, the more friction will be produced, tripling every five degrees as the inclination angle increases.7,32,33 There are also second and third order bends.8,13-39 Movement of the teeth while chewing can also be a source of friction at a given moment.14,22 In any case, the friction coefficient is greater for rectangular wires than for round wires and increases with wire thickness.9,29,30 It also increases with the use of esthetic materials (ceramic brackets).31 In addition, some researchers think saliva acts as a lubricant and decreases rubbing, while others consider that it can actually be an adhesion factor and therefore hinder sliding.1,28,29

The development of new self-proclaimed low friction or zero friction materials and techniques which promise to reduce length of treatment are based on changing the bracket design and the surface structure of the arch wire. This can be done with ion implantation,33-35,38 Poly (Chloro-P-Xylylene) coating, hot nitrogen diffusion,19 alumina crystalline, gas nitriding,36 carbon coating, and plasma38 and Teflon19 deposition. However, these changes in the arch wire structure could alter its fundamental properties, as in the case of Nickel-Titanium wire implanted with ions, which changes the elastic recovery of the wire.

Currently, a new compound is being studied called JAL 90458 (Orthospeed). Its colloid consistency as a gel coats surfaces that come into contact with it, neutralizing the forces of friction by acting as a buffer.39 The objective of this study was to measure the resistance to sliding according to the time and distance traveled of a stainless steel bracket on stainless steel arch wires of various sizes and cross sections with and without JAL 90458 on an inclined plane in order to prove that this compound reduces friction in the archwire-bracket coupling.

Material and Methodology

For the purposes of this study, we used a 35x40cm inclined plane (Figures 1 and 2) that has a 0.5cm groove in the middle where two rubber wheels can roll. One wheel goes in the base and the other in the mast, which will hold the wires at different angulations (45º and 60º) through little loops made of rigid 0.010" stainless steel ligatures attached to each wheel. On the wheel in the base, a small metal tube was placed, measuring 0.8cm long and 0.08cm in diameter. A channel measuring 0.04cm in diameter, which is held by rigid 0.010" ligature, will act as a guide for a 0.25mm-diameter and 50cm-long nylon line which will be used to pull the bracket.

Figure 1

Figure 1

Figure 2

Figure 2

Each wire was attached to the loops in the wheels made of 0.010" steel ligature wire. Another 0.010" steel wire loop was attached to each bracket with photopolymerized resin (Tetric® Ceram, Ivoclar Vivadent S.A. México). The 50-cm-long nylon line was tied to this loop in the bracket, then from there it was threaded through the small metal tube on the aforementioned wheel at the base and finally tied to some 150-gram weights. A protractor was adapted to act as a guide to measure the different angulations of the wires.

Six stainless steel 304VAR arch wires (DENSPLY GAC- Orthodontic, Lima, Peru) were used with the following characteristics: two 0.020" round arch wires, two 0.019x0.025" rectangular arch wires and two 0.021x0.025" rectangular arch wires. Each wire was 38cm long. Synergy® Rocky Mountain® straight wire low friction 0.022" slot size AISI 316 stainless steel brackets were used on the right side maxillary central incisors.

We used the following ligatures: 0.010", 5-cm-long, AISI 304VAR stainless steel ligature (DENSPLY GAC-Orthodontic), conventional 0.120" elastomeric ligatures (ORMCO) and 0.120" silicone injected low friction ligatures (Rocky Mountain®). Two types of JAL 90458 were used: Compound 1 was more fluid and Compound 2 had a more viscous consistency.39

Methodology

1.- The wires were placed at 45º and 60º degrees respectively in the measuring instrument starting with those wires of least caliber (0.020") and proceeding with the higher caliber wires (0.019x0.025" and 0.021x0.025"). The time and distance traveled by the bracket was recorded for each of the steel wires of differing calibers in combination with the three types of ligatures and angulations. All of this was done without using JAL 90458.

2.- JAL 90458 was applied immediately afterwards on the surface of the wires and brackets (compound 2) and on the ligatures (compound 1) and measurements of time and distance were taken.

We used 45º and 60º angulations due to the influence of the critical contact angle in second-order angulations. To ensure accurate measurements, a standard weight of 150 grams was used, since a lesser weight would not produce any movement, and a greater weight would make measuring the distance more difficult due to the speed at which the bracket would slide. One bracket for each wire caliber was used with and without the compound. The metal ligatures attached the wire with a total of 5 half turns for each test (procedure always carried out by the same person to avoid the bias of different ligation strengths). The time and distance measurements were single blind, in a dry environment at room temperature using a digital calibrator.

Determining Sliding Resistance

The maximum time for the bracket to slide the maximum distance from point A to point B along each wire was measured. The maximum time allotted was 30 seconds and the total distance was 30cm. If the bracket did not complete the 30-cm distance, a sliding time of 30 seconds was given.

Statistical Analysis

The data were analyzed using the statistical program SAS 9.1. The descriptive statistics of numeric independent variables for this test included the average and median of the time and distance of all the arch-bracket couples. The Kruskal-Wallis one-way analysis of variance by ranks for independent samples was applied to compare the different cross sections of wire, the different types of ligature and the different angulation. Statistical significance was (P ≤ .05).

Results

The tables we have included show the results obtained with and without JAL 90458 with the three different types of wire used in the study (0.20", 019.025" and 021.025"). Table I shows the results with and without the product comparing the three types of wire used with the three different types of ligatures (conventional elastic, low friction elastic and metal), all in relation to the time the bracket took to slide along the wire. We observed that without JAL the most time used was with the 021.025" wire and a conventional elastic ligature. With JAL 90458 the time used was much less. The comparison between the two groups (with and without JAL 90458) was highly significant.

Table I. Results with and without JAL 90458 of time slide to 60º of angulation.

Maximum time: 30 seconds. Minimum time: 0 seconds.
p: significance , *: p<0.05, **: p<0.01, ***: p<0.001, ns: no significant

Table II shows the same results but with a 45º angulation. As expected, the times were shorter at a greater angulation in the results without JAL 90458. However, when we applied the product, the results were not significantly different from those obtained at a 60º angulation. Nor were there differences with respect to the conventional elastic ligature which continues being the one with the greatest degree of sliding resistance. The differences between the two groups were also highly significant.

Table II. Results with and without JAL 90458 of time slide to 45º of angulation.

Maximum time: 30 seconds. Minimum time: 0 seconds.
p: significance , *: p<0.05, **: p<0.01, ***: p<0.001, ns: no significant

Table III shows the results obtained with and without JAL 90458 on conventional elastic ligatures in relation to the time the bracket took to move a certain distance. Without JAL 90458 the bracket did not slide at all. With the product, we observed that as the wire section increased the distance traveled decreased. Highly significant differences were found between the two groups.

Table III. Covered distance in the different types of wires with conventional elastic ligatures related with the time with and without JAL 90458, to 45º de angulation.

p: significance , *: p<0.05, **: p<0.01, ***: p<0.001, ns: no significant

When low friction elastic ligatures were used (Table IV), the bracket did move, even if JAL 90458 was not applied. In any case, the differences between the two groups were also statistically significant.<

Table IV. Covered distance in the different types of wires with elastic ligatures of low friction related with the time with and without JAL 90458, to 45º of angulation.

p: significance , *: p<0.05, **: p<0.01, ***: p<0.001, ns: no significant

Table V shows the results obtained when we used metal ligatures. The results are similar to those obtained with low friction elastic ligatures even though the distance traveled was greater with the metal ligatures compared to the low friction elastic ligatures.

Table V. Covered distance in the different types of wire with metallic ligatures related with the time with and without JAL 90458, to 45º of angulation.

p: significance , *: p<0.05, **: p<0.01, ***: p<0.001, ns: no significant

Discussion

In order to find solutions to sliding resistance, researchers are carrying out many different studies in the areas of bio-mechanics and mechanotherapy, thus favoring the development of new materials. The experiments done with techniques intended to reduce sliding resistance still entail problems.33,35,38 With JAL 90458 we find friction reduction to be very high, showing substantial statistically significant differences. This efficiency in reducing friction was shown for every type of ligature used. Some studies have not found differences regarding the types of ligature. Redlich36 did not find any differences with respect to the type of ligature when comparing conventional brackets to self-ligating brackets. We found the least friction using metal ligatures, followed by low friction elastic ligatures. This difference could be due to the method used to ligate the metal ligatures. Our results concur with those obtained by Fortín et al.37

Another important quality of this product is, in addition to reducing friction, it does not changes the properties of the wires.39 It only acts as a buffering agent between the wire and the bracket preventing rubbing between the two by sealing the rough surfaces. In contrast, ion implanted wires35 modify the springback quality of Ni-ti and affect the release of force of TMA® wires.

Most studies done on wire cross sections have found that the bigger the wire is, the more friction is produced.5, 7, 11, 12, 14, 19, 20, 22, 26, 27, 34 Our study coincides with these results. This difference was most evident when JAL 90458 was not applied. However, when we used the product, the differences found between wire sizes were not so pronounced.

Conclusions


JAL 90458 significantly reduced sliding resistance on any stainless steel wire-bracket couple.
JAL 90458 significantly reduced friction on all the ligatures studied.
The cross section and caliber of the wires did not have a great influence on friction increase when JAL 90458 was used compared to the results obtained in the group that did not use the product.
When JAL 90458 was not used, the metal ligature created the least amount of friction.
The wire-bracket couple sliding time when JAL 90458 was used was always less than when the compound was not used. The distance traveled by the brackets was always greater when the JAL 90458 compound was used than when the test was carried out without applying the compound.

IProfessor and Chair Orthodontic Department, Universidad Complutense de Madrid, Madrid, Spain.
IIMaster of Dental Science, Orthodontic Department, Universidad Complutense de Madrid, Madrid, Spain.
IIIAssistant Professor European Institute of Technology for Dental Science, Universidad de Alcalá, Alcalá de Henares, Madrid, Spain.
IVProfessor of Master in Orthodontics, Universidad de Alcalá,, Madrid, Spain.
VMedical Director of Institute of Investigation in Orthodontics.

Corresponding Author: Dr. Juan J. Alió. Facultad de Odontología. Plaza Ramón y Cajal s/n. Universidad Complutense de Madrid. 28040 Madrid. ( Esta dirección de correo electrónico está siendo protegida contra los robots de spam. Necesita tener JavaScript habilitado para poder verlo. ).

Referance

1. Taylor NG. Ison K. Frictional resistance between orthodontic brackets and archwires in the bucal segments. Angle Orthod. 1995;66:215-222.
2. Wadhwa PK. Kwon HK. Friction resistance in ceramic and metal brackets. J Clin Orthod. 2004;38:35-38.
3. Yukio K, Hisao F. Numeric simulations of en-masse space closure with sliding mechanics. Am J Ortho Dentofac Orthop. 2010;138:704-710.
4. Staggers JA, Germane N. Clinical considerations in the use of retraction mechanics. J Clin Orthod. 1991; 25(6): 364-369.
5. Tselepsis M. The dynamic frictional resistance between orthodontic brackets and archwires. Am J Orthod Dentofacial Orthop 1994;106:131-138.
6. Rossouwn PE. Friction: An Overview. Seminars in Orthodontics 2003;9(4):218-222.
7. Matarese G, Nucera R, Militi A. Evaluation of frictional forces during dental alignment: An experimental model with 3 nonleveled brackets. Am J Orthod Dentofacial Orthop. 2008;133:708-715.
8. Krim J. Surface science and the atomic-scale origins of friction: what once was old is new again. Surface Science. 2002;500:741-758.
9. Rossouwn E. A fundamental review of variables associated with low velocity frictional dynamics. Seminars orthod. 2003;9(4):223- 235.
10. Lorenz Braunchi. Chirstiane Senn. Active and passive self-ligation - a myth? Angle Orthod. 2011;81:312-318.
11. Kusy RP. Whitley JQ. Friction between different wire-bracket configurations and materials. Semin Orthod 1997;3:166-177.
12. Whitley J. Kusy RP. Influence of interbracket distances on the resistance to sliding of orthodontic appliances. Am J Orthod Dentofacial Orthop. 2007;132:360-362.
13. Bo-Sun K. Seung-Hak B. Three-dimensional relationship between the critical contact angle and the torque angle. Am J Orthod Dentofacial Orthop. 2003;123:64-73.
14. Kusy RP, Withley JQ. Influence of archwire and brackets dimensions on sliding mechanics: derivation and determinations of the critical contact angles for binding. Eur J Orthod 1999;21:199-208.
15. Burrow JS. Friction and resistance to sliding in orthodontics: A critical review. Am J Orthod Dentofacial Orthop. 2009;135:422-447.
16. Swartz ML. Fact or friction: The clinical relevance of in Vitro steady-state friction studies. J Clin Orthod. 2007;41(8):427-431.
17. Henao S. Fricctional evaluations of dental typodonts models using four self-ligation designs and a conventional designs. Angle Orthod. 2004;75:75-85.
18. Suárez D. Problemas tribológicos en el diseño de brackets. Rev Esp ortod. 1995;25:29-45.
19. Hussmann P, Bourauel C, Wessinger M, Jager A. The frictional behavior of coated guiding archwires. J Orofac Orthop. 2002;63:199-211.
20. Hamdan Ahmad. The effect of different combinations of tip and torque on archwire/bracket friction. Eur J Orthod. 2008;30:508-514.
21. Hain M, Dhopatkar A, Rock P. The effect of ligation method on friction in sliding mechanics. Am J Orthod Dentofacial Orthop. 2003;123:416-22.
22. Ireland A, Mc Donald F. Effect of bracket and wire composition on frictional forces. Eur J Orthod. 1991;13:322-28.
23. Bobadilla M. Medición in vitro de la fuerza de fricción en duplas arco-bracket con angulación. Rev Ing Biomed. 2008;2(3):84-90.
24. Fisher H, Souni M. Transformation behavior, chemical composition, surface topography and bending properties of five selected 0.016 x 0.022 NiTi archwires. J Orofac Orthop. 2003;64:88-99.
25. Bourauel C, Fries T, Drescher D. Surface roughness of orthodontics wires via atomic force microscopy, laser reflectance, and profilometry. Eur J Orthod. 1998;20:79-92.
26. Ogata R, Nanda R, Currier F. Frictional resistances in stainless steel bracket-wire combinations with effects of vertical deflections. Am J Orthod Dentofacial Orthop. 1996;109:535-42.
27. Kapila S, Angolcar P, Duncanson M. Evaluation of friction between edgewise stainless steel brackets and orthodontic wires of four alloys. Am J Orthod Dentofacial Orthop. 1990;98:117-26.
28. Thorstenson GA, Kusy RP. Resistance to sliding of self-ligating brackets versus conventional stainless steel twin brackets with second-order angulations in the dry and wet states. Am J Orthod Dentofacial Orthop. 2001;120:361-70.
29. Kusy R. Comparison of the frictional coefficients for selected archwire-bracket slot combinations in the dry and wet states. Angle Orthod. 1991;61(4):293-302.
30. Chin Liang Yeh. In vitro evaluations of fricctional resistance between brackets with passive-ligation designs. Am J Orthod Dentofacial Orthop. 2007;131:704.e11-704.e22.
31. Kusy R, Whitley J. Comparisons of surface roughnesses and sliding resistances of 6 titanium-based or TMA-type archwires. Am J Orthod Dentofacial Orthop. 2004;126:589-603.
32. Moore M. Factors affecting friction in the pre-adjusted appliance. Eur J Orthod.. 2004;26:579-583.
33. Chin Liang Yeh. In vitro evaluations of fricctional resistance between brackets with passive-ligation designs. Am J Orthod Dentofacial Orthop 2007;131:704.e11-704.e22.
34. Kula K. Effect of implantation of TMA archwires on the rate of orthodontic sliding space closure. Am J Ortho Dentofacial Orthop. 1998;114:577-81.
35. Ryan R, Walker G, Freeman K, Cisneros G. The effects of ion implantation on rate of tooth movement: An in vitro model. Am J Orthod Dentofacial Orthop 1997;112(1):64-68.
36. Redlich M. el al. In Vitro study of frictional forces during sliding mechanics of reduced friction brackets. Am J Orthod Dentofac Orthop 2003, 1: 69-73.
37. Fortini A, Lupoli M, Cacciafesta V. A new low friction ligation system. J Clin Orthod 2005, 8: 434-470.
38. Zufall. Sliding mechanics of coated composites wires and the development of an engineering model for binding. Angle Orthod 2000;70:34-47.
39. Alió Sanz en Expoorto 09: Análisis de fiabilidad y validación científica en la fase de nivelación del JAL 90458 como producto que disminuye la fricción en el tratamiento con aparatos fijos multibrackets. 1a ed. Madrid: Ripano S.A.; 2009. p. 195-201.